Bovine IgG1 (Immunoglobulin G1) ELISA Kit

Bovine IgG1 (Immunoglobulin G1) ELISA Kit

Size: 96T Price: $480
Reactive Species: Bovine
Assay Range: 61.7-5000ng/mL
Sensitivity: 29.1ng/mL
SKU: XEB1044 Category: Bovine Tags: ,

Datasheet

INFORMATION

Full NameImmunoglobulin G1
SynonymIGHG1, Ig Gamma-1 Chain C Region, Immunoglobulin Heavy Constant Gamma 1, G1m Marker, Immunoglobulin Gm1
Assay TypeCompetitive
ReactivityBovine
Range61.7-5000ng/mL
Sensitivity29.1ng/mL
Sample TypeSerum, plasma and other biological fluids
Sample Volume50μL
Detection WavelengthOD450
Reaction Duration1.5h
Transportation Temperature2-8°C
SpecificitySpecifically binds with IgG1, no obvious cross reaction with other analogues.
Microplate96-wells plate breakable into 12 x 8 wells strip

ELISA KIT COMPONENTS

Upon receipt, unpack promptly and store as recommended in the instructions.

ComponentsSpecificationsStorage and Notes
Micro Plate
96T: 8 wells×12 strips
48T: 8 wells×6 strips
Unopened: -20°C, 12 months
Unused: Put it back in the aluminum foil bag and seal it, store it at -20°C.
Reference Standard
96T: 2 vials
48T: 1 vial
Unopened: -20°C, 12 months
Please use freshly dissolved standards for each experiment. Discard any unused standards after dissolution.
Biotinylated Detection Ab Concentrate (100×)
96T: 120μL×1 vial
48T: 60μL×1 vial
Unopened: -20°C, 12 months
Unused: Please seal the concentrate and store it at -20°C, and discard the working solution.
HRP Conjugate Concentrate (100×)
96T: 120μL×1 vial
48T: 60μL×1 vial
Unopened: -20°C(Protect from light), 12 months
Unused: Please seal the concentrate and store it at -20°C, and discard the working solution.
Biotinylated Detection Ab Diluent
14mL×1
2-8℃, 12 months
HRP Conjugate Diluent
14mL×1
2-8℃, 12 months
Reference Standard & Sample Diluent20mL×1
2-8℃, 12 months
Washing Buffer Concentrate (25×)30mL×1
2-8℃, 12 months
Substrate Reagent(TMB)10mL×12-8°C(Protect from light),12 months
Stop Solution7mL×12-8°C/Room temperature

REQUIRED INSTRUMENTS AND REAGENTS

  • Microplate reader (450nm wavelength filter).
  • Incubator capable of maintaining 37℃.
  • 1.5mL EP tube, absorbent paper.
  • Pipette and disposable tip: 0.5-10μL, 2-20μL, 20-200μL, 200-1000μL.
  • Double distilled water or deionized water.

ASSAY PROCEDURE SUMMARY

1
Add 50μL standard or sample to each well, add 50μL Biotinylated Detection Ab working solution to each well. Incubate at 37°C for 45 min.
Washing: Aspirate and wash the plate for 3 times.
2
Add 100μL HRP conjugate working solution.Incubate at 37°C for 30 min.
Washing: Aspirate and wash the plate for 5 times.
3
Add 90μL Substrate Reagent. Incubate at 37℃ for 15 min.
4
Add 50μL Stop Solution.
5
Readings at 450nm. Calculation of results.

STANDARD CURVE

The product has been verified by the quality inspection department and meets the performance requirements of the product manual. (Laboratory temperature: 20-25℃, laboratory relative humidity: 20-60%. Before the experiment, all components were equilibrated to room temperature.) The following data is for reference only. A standard curve must be run with each test.
STD.(ng/mL)OD-1OD-2Average
02.2812.2972.289
78.131.9741.9761.975
156.251.6071.6051.606
312.51.2411.2231.232
6250.8670.8670.867
12500.6090.5770.593
25000.4130.4370.425
50000.3170.3530.335
 Standard Curve Image

RECOVERY

The recovery of Bovine Immunoglobulin G1 spiked to three different levels in samples throughout the range of the assay in various matrices was evaluated.

LINEARITY

To assess the linearity of the assay, samples containing and/or spiked with high concentrations of Bovine Immunoglobulin G1 in various matrices were diluted with the Reference Standard & Sample Diluent to produce samples with values within the dynamic range of the assay.

PRECISION(%)

Intra-Assay Precision (Precision within an assay): Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-Assay Precision (Precision between assays): Three samples of known concentration were tested in twenty separate assays to assess inter-assay precision. Assays were performed by at least three technicians using two lots of components.

ItemIntra-assay PrecisionInter-assay Precision
Sample123123
n202020202020
Mean (ng/mL)228.85595.412168.6218.86576.882253.85
Standard deviation14.2630.6697.3714.5532.3197.14
CV(%)6.235.154.496.655.64.31

STABILITY

Perform stability testing of the kit at 37°C and 2-8°C and obtain relevant data.

ELISA kit(n=5)37°C for 1 month2-8°C for 6 months
Average(%)8095-100