INFORMATION
Full Name | Albumin |
---|---|
Synonym | ALB |
Assay Type | Sandwich ELISA |
Reactivity | Chicken |
Range | 3.13-200μg/mL |
Sensitivity | 1.88μg/mL |
Sample Type | Serum, plasma and other biological fluids |
Sample Volume | 100μL |
Detection Wavelength | OD450 |
Reaction Duration | 3.25h |
Transportation Temperature | 2-8°C |
Specificity | Specifically binds with ALB, no obvious cross reaction with other analogues. |
Microplate | 96-wells plate breakable into 12 x 8 wells strip |
ELISA KIT COMPONENTS
Upon receipt, unpack promptly and store as recommended in the instructions.
Components | Specifications | Storage and Notes |
---|---|---|
Micro Plate | 96T: 8 wells×12 strips 48T: 8 wells×6 strips | Unopened: -20°C, 12 months Unused: Put it back in the aluminum foil bag and seal it, store it at -20°C. |
Reference Standard | 96T: 2 vials 48T: 1 vial | Unopened: -20°C, 12 months Please use freshly dissolved standards for each experiment. Discard any unused standards after dissolution. |
Biotinylated Detection Ab Concentrate (100×) | 96T: 120μL×1 vial 48T: 60μL×1 vial | Unopened: -20°C, 12 months Unused: Please seal the concentrate and store it at -20°C, and discard the working solution. |
HRP Conjugate Concentrate (100×) | 96T: 120μL×1 vial 48T: 60μL×1 vial | Unopened: -20°C(Protect from light), 12 months Unused: Please seal the concentrate and store it at -20°C, and discard the working solution. |
Biotinylated Detection Ab Diluent | 14mL×1 | 2-8℃, 12 months |
HRP Conjugate Diluent | 14mL×1 | 2-8℃, 12 months |
Reference Standard & Sample Diluent | 20mL×1 | 2-8℃, 12 months |
Washing Buffer Concentrate (25×) | 30mL×1 | 2-8℃, 12 months |
Substrate Reagent(TMB) | 10mL×1 | 2-8°C(Protect from light),12 months |
Stop Solution | 7mL×1 | 2-8°C/Room temperature |
REQUIRED INSTRUMENTS AND REAGENTS
- Microplate reader (450nm wavelength filter).
- Incubator capable of maintaining 37℃.
- 1.5mL EP tube, absorbent paper.
- Pipette and disposable tip: 0.5-10μL, 2-20μL, 20-200μL, 200-1000μL.
- Double distilled water or deionized water.
ASSAY PROCEDURE SUMMARY
STANDARD CURVE
STD.(μg/mL) | OD-1 | OD-2 | Average |
---|---|---|---|
0 | 0.06 | 0.072 | 0.066 |
3.13 | 0.109 | 0.117 | 0.113 |
6.25 | 0.169 | 0.141 | 0.155 |
12.5 | 0.231 | 0.207 | 0.219 |
25 | 0.366 | 0.37 | 0.368 |
50 | 0.893 | 0.867 | 0.88 |
100 | 1.502 | 1.51 | 1.506 |
200 | 2.313 | 2.315 | 2.314 |

RECOVERY
5 different levels of target protein spiked in samples were evaluated for recovery experiments to determine the range of recovery and calculate the average recovery rate.
LINEARITY
5 different levels of target protein spiked in samples were evaluated for recovery experiments to determine the range of recovery and calculate the average recovery rate. Subsequently, the 5 samples were diluted to 2-fold, 4-fold, 8-fold, and 16-fold 10 concentrations, and recovery experiments were performed again to evaluate the range and average recovery rate at different dilution factors.
PRECISION(%)
Intra-assay precision (Precision within an assay): Low, mid range and high levels of the samples were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): Low, mid range and high levels of the samples were tested on 3 different plates, 20 replicates in each plate, respectively.
Item | Intra-assay Precision | Inter-assay Precision | ||||
---|---|---|---|---|---|---|
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (μg/mL) | 10.26 | 25.95 | 73.37 | 10.3 | 27.35 | 80.18 |
Standard deviation | 0.63 | 1.28 | 2.98 | 0.66 | 1.58 | 3.01 |
CV(%) | 6.13 | 4.92 | 4.06 | 6.41 | 5.77 | 3.76 |
STABILITY
Perform stability testing of the kit at 37°C and 2-8°C and obtain relevant data.
ELISA kit(n=5) | 37°C for 1 month | 2-8°C for 6 months |
---|---|---|
Average(%) | 80 | 95-100 |