Guinea pig MUC5AC (Mucin 5 Subtype AC) ELISA Kit

Guinea pig MUC5AC (Mucin 5 Subtype AC) ELISA Kit

Size: 96T Price: $480
Reactive Species: Guinea pig
Assay Range: 31.25-2000pg/mL
Sensitivity: 18.75pg/mL
SKU: XEGP1350 Category: Guinea pig Tags: ,

Datasheet

INFORMATION

Full NameMucin 5 Subtype AC
SynonymMUC5AC
Assay TypeSandwich ELISA
ReactivityGuinea pig
Range31.25-2000pg/mL
Sensitivity18.75pg/mL
Sample TypeSerum, plasma and other biological fluids
Sample Volume100μL
Detection WavelengthOD450
Reaction Duration3.25h
Transportation Temperature2-8°C
SpecificitySpecifically binds with MUC5AC, no obvious cross reaction with other analogues.
Microplate96-wells plate breakable into 12 x 8 wells strip

ELISA KIT COMPONENTS

Upon receipt, unpack promptly and store as recommended in the instructions.

ComponentsSpecificationsStorage and Notes
Micro Plate
96T: 8 wells×12 strips
48T: 8 wells×6 strips
Unopened: -20°C, 12 months
Unused: Put it back in the aluminum foil bag and seal it, store it at -20°C.
Reference Standard
96T: 2 vials
48T: 1 vial
Unopened: -20°C, 12 months
Please use freshly dissolved standards for each experiment. Discard any unused standards after dissolution.
Biotinylated Detection Ab Concentrate (100×)
96T: 120μL×1 vial
48T: 60μL×1 vial
Unopened: -20°C, 12 months
Unused: Please seal the concentrate and store it at -20°C, and discard the working solution.
HRP Conjugate Concentrate (100×)
96T: 120μL×1 vial
48T: 60μL×1 vial
Unopened: -20°C(Protect from light), 12 months
Unused: Please seal the concentrate and store it at -20°C, and discard the working solution.
Biotinylated Detection Ab Diluent
14mL×1
2-8℃, 12 months
HRP Conjugate Diluent
14mL×1
2-8℃, 12 months
Reference Standard & Sample Diluent20mL×1
2-8℃, 12 months
Washing Buffer Concentrate (25×)30mL×1
2-8℃, 12 months
Substrate Reagent(TMB)10mL×12-8°C(Protect from light),12 months
Stop Solution7mL×12-8°C/Room temperature

REQUIRED INSTRUMENTS AND REAGENTS

  • Microplate reader (450nm wavelength filter).
  • Incubator capable of maintaining 37℃.
  • 1.5mL EP tube, absorbent paper.
  • Pipette and disposable tip: 0.5-10μL, 2-20μL, 20-200μL, 200-1000μL.
  • Double distilled water or deionized water.

ASSAY PROCEDURE SUMMARY

1
Add 100μL standard or sample to each well, incubate at 37℃ for 90 min. Discard the liquid.
2
Add 100μL Biotinlated Detection Ab working solution to each well. Incubate at 37℃ for 60 min.
Washing: Aspirate and wash the plate for 3 times.
3
Add 100μL HRP conjugate working solution. Incubate at 37℃ for 30 min.
Washing: Aspirate and wash the plate for 5 times.
4
Add 90μL Substrate Reagent. Incubate at 37℃ for 15 min.
5
Add 50μL Stop Solution.
6
Readings at 450nm. Calculation of results.

STANDARD CURVE

The product has been verified by the quality inspection department and meets the performance requirements of the product manual. (Laboratory temperature: 20-25℃, laboratory relative humidity: 20-60%. Before the experiment, all components were equilibrated to room temperature.) The following data is for reference only. A standard curve must be run with each test.
STD.(pg/mL)OD-1OD-2Average
00.0890.0890.089
31.250.1360.1460.141
62.50.1950.1850.19
1250.3110.2870.299
2500.5230.5610.542
5000.9860.980.983
10001.6851.7331.709
20002.5292.5312.53
 Standard Curve Image

RECOVERY

5 different levels of target protein spiked in samples were evaluated for recovery experiments to determine the range of recovery and calculate the average recovery rate.

LINEARITY

5 different levels of target protein spiked in samples were evaluated for recovery experiments to determine the range of recovery and calculate the average recovery rate. Subsequently, the 5 samples were diluted to 2-fold, 4-fold, 8-fold, and 16-fold 10 concentrations, and recovery experiments were performed again to evaluate the range and average recovery rate at different dilution factors.

PRECISION(%)

Intra-assay precision (Precision within an assay): Low, mid range and high levels of the samples were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): Low, mid range and high levels of the samples were tested on 3 different plates, 20 replicates in each plate, respectively.

ItemIntra-assay PrecisionInter-assay Precision
Sample123123
n202020202020
Mean (pg/mL)98.63310.01748.24100.52308.5701.88
Standard deviation5.217.8936.665.2314.7231.44
CV(%)5.275.774.95.24.774.48

STABILITY

Perform stability testing of the kit at 37°C and 2-8°C and obtain relevant data.

ELISA kit(n=5)37°C for 1 month2-8°C for 6 months
Average(%)8095-100